Milli-Q® IQ 7000 specifications

Specifications

Feed water requirements
Feed water quality Pretreated water using EDI, DI, RO or distillation technologies
Conductivity at 25 °C < 100 µS/cm
Total Organic Carbon (TOC) < 50 ppb
Pressure range * Max 6 bar
Temperature 5 – 35 °C / 41 – 95 °F
* For pressures between 0 and 0.1 bar, the system will operate, but product flow rate may be lower.

Ultrapure, Type 1 water specifications1 (from Q-POD® dispenser)
Resistivity at 25 °C2 18. 2 MΩ·cm
TOC3 ≤ 2 ppb (µg/L)
Particles4 No particles with size > 0.22 µm
Bacteria < 0.01 CFU/mL4,5
< 0.005 CFU/mL6
Pyrogens (endotoxins)5 < 0.001 EU/mL
RNases5 < 1 pg/mL
DNases5 < 5 pg/mL
Proteases5 < 0.15 µg/mL
Flow rate 0.05 – 2 L/min
These values are typical and may vary depending on the nature and concentration of contaminants in the feed water.
2 Resistivity can also be displayed non-temperature-compensated as required by USP.
In the appropriate operating conditions.
With Millipak® and Millipak® Gold
With Biopak®
With Millipak® Gold installed and used in a laminar flow hood

International regulation requirements

Declaration of CE and cUL Conformity:
The Milli-Q® IQ 7000 system has been tested by an independent and accredited company for compliance with CE directives related to safety and electromagnetic compatibility. The report can be consulted on demand
. In addition, the Milli-Q® IQ 7000 system is built using components and practices recommended by UL and has been cUL marked